Luminometer |
Orion II Technical Data |
Sample Format |
96-well microplate, opaque in solid or strip format, |
Detector |
Photomultiplier tube with bialkali cathode, effective spectral sensitivity |
Sensitivity |
2 atto mole ATP. |
Crosstalk |
Less than 3 x 10-5. |
Dynamic Range |
6 decades. |
Measuring Time |
0.1-100 s per well. |
Scan Pattern |
Measurement by row or by columns. |
Injectors |
Up to four per Orion II system (upgradeable). |
Injection Volumes |
10-150 ¥ìl per pump in increments of 1¥ìl. |
Injection Pattern |
Double Injection Design DID: Simultaneous injection into two adjacent |
Tubing |
Chemically inert PTFE tubing and connections (PTFE; KEL-F; glass; PS), |
Priming |
Forward and reverse priming, external priming container or built-in priming |
Shaker |
Linear, orbital and cross shaking. |
Sample Incubation |
Ranges for optional sample incubation: |
Automation |
Compatible to all known microplate transfer devices. |
Interfaces |
Serial interface (RS-232) and USB interface for PC. |
Dimensions |
W 385 mm, D 410 mm, H 255 mm. |
Weight |
22 kg. |
Power |
230 V 50 Hz; 115 V 60 Hz. |
Power Consumption |
30 VA, maximum 70 VA (4 injectors and heating). |